IMPORTANT
!PLEASE ENSURE YOU CONTACT BEAMLINE STAFF, IF YOU HAVE NOT ALREADY, TO DISCUSS USE OF THE REIXS-SPA GLOVEBOX AS IT REQUIRES AT LEAST 1-DAY NOTICE TO PURGE THE GLOVEBOX PRIOR TO USE!
General
The argon flow glovebox will be setup prior to commencement of your experiment. Please do not adjust the flow rate of close/open any valves that are covered in the user instructions below.
Inserting Samples/Equipment
Samples and equipment can inserted through the attached antechamber. There are two methods in which items can inserted into the glove box:
-
The antechamber can be purged with Ar gas; all air is pushed out the check-valve at the top of the chamber. This is required for placing items that can not handle vacuum such as liquids and fragile air filled bags. This takes approximately 30 minutes and it may be there is still trace air/moisture that is trapped. When these items are inserted into the glovebox, let them acclimate further prior to commencing work for ~30 minutes to ensure all air and moisture is removed.
-
The antechamber can be pump/purged with Ar gas. This requires the evacuation and letup of the antechamber with Ar gas at least 3 times to ensure no moisture of air remains. This is much more effective that the passive purge method, but requires vacuum compatible items. This method takes approximately 30 minutes to complete.
Antechamber Passive Purge
The following steps should be followed when passively purging the antechamber and inserting samples/equipment into the REIXS-SPA Ar-Flow Glovebox.
-
Open the antechamber door.
-
Fully loosen centre knob.
-
Lift securing latch.
-
Open door fully.
-
-
Places samples into the antechamber.
-
Ensure all your samples/equipment is oriented to allow air to rise.
-
Use the SS tray provided.
-
This will make moving the samples/equipment into the glovebox much easier.
-
-
-
Ensure the O2 meter is installed in the antechamber.
-
Check with beamline staff if no O2 monitor is present.
-
-
Close the antechamber door.
-
Lift the securing latch.
-
Tighten the central knob hand-tighten.
-
This will bottom out, so there is no benefit to overtightening.
-
-
-
Open the argon supply knob fully.
-
Check the flow rate is ~10 cph.
-
-
Allow the chamber to purge until 0.0% O2 is recorded.
-
This will take ~30 minutes.
-
Antechamber Pump/Purge
The following steps should be followed when actively pump/purging the antechamber and inserting samples/equipment into the REIXS-SPA Ar-Flow Glovebox.
-
Open the antechamber door.
-
Fully loosen centre knob.
-
Lift securing latch.
-
Open door fully.
-
-
Places samples into the antechamber.
-
Ensure all your samples/equipment is oriented to allow air to rise.
-
This is not required, but helps with cleaning the environment.
-
-
Use the SS tray provided.
-
This will make moving the samples/equipment into the glovebox much easier.
-
-
Ensure your samples/equipment are vacuum compatible.
-
-
Ensure O2 meter is not installed in the antechamber.
-
Check with beamline staff if O2 monitor is present.
-
-
Close the antechamber door.
-
Lift the securing latch.
-
Tighten the central knob hand-tighten.
-
This will bottom out, so there is no benefit to overtightening.
-
-
-
Open the valve to the vacuum pump.
-
This will be the valve without the flowmeter farthest away.
-
-
Allow the pump to evacuate the antechamber to to less than 26 inHg vacuum.
-
Achieving fully vacuum (29.9 inHg) will take excessive time and repetitive pump/purging is more efficient.
-
-
Close the valve to the vacuum pump and open the valve to the argon gas.
-
The supply valve will have the flowmeter attached.
-
-
Allow the pressure to rise to 0 psig (relative to atmospheric pressure).
-
This will take several minutes.
-
No need to watch closely as there is a PRV to prevent overpressure.
-
-
Repeat steps 5-8 two more times to ensure a water and oxygen free environment before inserting samples/equipment into the glovebox.
Working in the Glovebox
Prior to, and after, samples/equipment are inserted into the glovebox, one should familiarize oneself with the use of the glovebox. While there are no strict procedures when working in the glovebox, below are some helpful tips that should be followed to ensure incident free use of the REIXS-SPA Ar-Flow Glovebox:
-
Ensure you wear a long sleeve clothing and don nitrile gloves.
-
The glovebox is operated with a constant flow of 2cph of argon. This is sufficient to ensure replenishment of argon that expelled during normal use. However, care should be taken when inserting or removing your hands from the gloves.
-
Insert your hands slowly and try not insert the gloves further than required.
-
Using the gloves, from the side, pull the gloves onto your hands.
-
When removing your hands, use the gloves to carefully pull the gloves off without fully withdrawing the gloves (this is atypical where a valve will open to quickly refill the glovebox). This will ensure you do not create a vacuum and draw air into the glovebox.
-
-
Don additional gloves on the glovebox built-in gloves. This will ensure manual dexterity is improved.
-
Do not be gentle when pulling on additional gloves.
-
-
Achieve a comfortable stance and ensure you can comfortably operate the glovebox.
-
Discuss with beamline staff if you require adjustment of the platform.
-
-
Ensure you close the antechamber door when finished with sample/equipment insertion.
-
Accidental contamination of the glovebox may require up to 24hrs to restore an oxygen free environment.
-
-
When working with sharp tools in the glovebox, take extra care not to damage the glovebox gloves.
Mounting Samples
Powder Samples
Powder samples are problematic for mounting in a glovebox due the increased risk of contamination and waste generation. At the present time we do not allow toxic powders to be mounted in the glove box. Pre-mounted powder samples at the user host institution are allowed as trace waste generation will be minimal.
Below are the policies that shall be followed when mounted powder samples in the REIXS-SPA Ar-Flow Glovebox:
-
Only use a minimal amount of powder sample.
-
Clean up excess powder with wetted wipe (isopropanol).
-
Ensure you place waste in the bag provided.
-
-
Use wetted cotton swab to clean excess powder from sample plates.
Below are the general steps that should be followed when mounting powder samples:
-
Prepare mounting medium on the sample plate.
-
Generally this can be done outside the glovebox, which will be much easier.
-
-
Dispense small amount of powder onto the mounting medium.
-
Work on the large SS tray using a clean wipe to contain excess contaminating powder.
-
-
Close main sample container.
-
Work powder into mounting medium with spatula or equivalent tool.
-
Clean spatula and set aside.
-
Remove excess powder by tapping upside down such that the excess powder does not cross existing samples.
-
Ensure clean wipe is on top of the tapping block.
-
Make strong taps to ensure excess powder is removed.
-
-
Repeat the preceding steps for additional samples.
Solid/Pre-mounted Samples
Mounting solid or pre-mounted samples requires no additional instructions. Generally, these materials are mounted using carbon tape. Just remember that excess carbon will create challenges in UHV environments.
Inserting/Removing Samples into the Transfer Cart
Currently carried out by beamline staff. Users instructions will be provided once an appropriate procedure is determined.
Inserting/Removing Samples into the Beamline Loadlock
Currently carried out by beamline staff. Users instructions will be provided once an appropriate procedure is determined.
Transferring Samples to the Beamline
Currently, operation of the transfer cart must be completed by beamline staff. Inserting/removal of samples into transfer cart sample plate holder can be completed by users, see above, but all movement/coupling of the transfer cart to the glovebox and beamline must be completed by beamline staff. This is due to the large mass of the transfer cart that carries significant risk of damage to equipment and persons without appropriate PPE.
Prohibited Items
In order to ensure continued operation of the REIXS-SPA Ar-Flow Glovebox for all users, the following items may not be inserted/used:
-
Silver Paint
-
Solvents; other than isopropanol
-
Excessively heavy objects
-
Toxic powders; beamline staff will make the final decision
-
Flammable/oxidizing materials